1. Baseline conditions and working conditions
Influence quantity
Datum condition
Working conditions
Remarks
ambient temperature
23 C + 1 c
- 10 ~40 C
Environmental wet crossing
40% to 60%
< 80%
Signal waveform
sine wave
sine wave
Beta =0.05
signal frequency
50HZ + 1HZ
45H ~ 65HZ
Instrument operating voltage
9V + 0.1V
9V + 1V
Measuring Phase Current Amplitude
1A + 0.1A
10m ~ 20.0A
Phase Voltage Amplitude Measurement
100V + 10V
30V ~ 600V
External electric field and magnetic field
Should avoid
Location of measured conductor
Approximate geometric center position of the measured wire at the clamp
2. General specification
Function
Direct testing of phase, AC current, AC voltage, AC leakage current and phase sequence; identification of transformer groups, inductive and capacitive circuits; indirect testing of power factor and power
Power Supply
DC9V alkaline dry cell (1.5V AA*6)
Power waste
Turn on the backlight for about 35 mA and the battery works for about 40 hours continuously.
Turn off the backlight, the instrument consumes about 5mA, and the battery works for about 300 hours continuously.
display mode
LCD display, blue screen backlight function, suitable for dark places
LCD size
70mm * 62mm
LCD display domain
64mm * 54mm
Instrument size
Length, width and thickness: 207mm *101mm *45mm
Jaw size
_7.5 mm *13 mm
Sampling rate
About 3 times / sec.
Range
AC Voltage: 0-20V/200V/600V
AC current: 0-200 mA/2A/20A
Phase: 0-360 degrees
Phase measurement
Signal under test
Amplitude range
When measuring the phase of U1-U2: 30V-600V
Measuring I1-I2 phase: 10mA-20.00A
When measuring the phase of U1-I2 or I1-U2: 10V-600V, 10mA-20.00A
Data retention
Press HOLD key to keep data in test, and "DH" symbol display
Automatic shutdown
After about 15 minutes, the instrument shuts down automatically to reduce battery consumption.
Voltage detection
When the battery voltage is less than 7.2V, the low symbol of the battery voltage is displayed to remind the battery to be replaced.
Instrument quality
The main engine is about 550G (with batteries), the meter clamp is about 170g *2, and the test line is about 250g.
Test line length
1.5m
Current clamp length
2m
Working temperature and humidity
- 10 ~40 below 80% Rh
Storage temperature and humidity
- 10 ~60 below 70% Rh
Input impedance
The input impedance of each test voltage is: 2M
Voltage input impedance of U1U2 phase measurement is 40K
Withstand voltage
Sinusoidal AC voltage withstanding 2000V/50Hz between instrument top cover and battery cover screw lasts 1 minute
Insulation
Instrument Circuit (> 10M) Between Outer Shell and Two Voltage Input Terminals
Structure
Double insulation
3. Basic Errors and Performance Indicators under Baseline Conditions
Category
Range
Resolving power
Fundamental error
Voltage
20V
0.01V
(1.2% RDG + 2dgt)
200V
0.1V
500V
1V
Electric current
200mA
0.1mA
(1.0% RDG + 2dgt)
2A
1mA
10A
10mA
Phase
0~360 degrees
1 degrees
+ 1 degrees
Note: Under working conditions, the phase error does not exceed 3 degrees.