1. Baseline conditions and working conditions
Influence quantity
Datum condition
Working conditions
Remarks
ambient temperature
23 C + 1 c
- 10 ~40 C
/
Environmental wet crossing
40% to 60%
< 80%
/
Signal waveform
sine wave
sine wave
Beta =0.01
signal frequency
50HZ + 1HZ
45HZ ~ 65HZ
/
Instrument operating voltage
3.7V + 0.1V
3.7V + 0.5V
/
Phase Frequency Permissible Current Amplitude
1A + 0.1A
1mA ~ 20.0A
/
Voltage Amplitude in Phase Frequency Phase Sequence Measurement
100V + 10V
10V ~ 600V
/
Current Amplitude in Measuring Power Factor
1A + 0.1A
10mA ~ 20A
/
Voltage Amplitude in Measuring Power Factor
100V + 10V
10V ~ 600V
/
External electric field and magnetic field
Should avoid
Location of measured conductor
Approximate geometric center position of the measured wire at the clamp
2. General specification
Function
At the same time, the current, voltage, phase, frequency, phase sequence, active power, reactive power, apparent power, power factor, current vector sum of three-phase 380V/220V power system are tested. The transformer wiring groups, inductive and capacitive circuits are distinguished. The secondary circuit and bus differential protection system are tested. The phase relations between CT groups of differential protection are read out, and the correct wiring of the watt-hour meter is checked. No, overhaul line equipment, etc.
Power Supply
DC 3.7V large capacity rechargeable lithium battery, USB charging interface, full charge continuous work for about 10 hours
Rated current
About 150mA
display mode
3.5 inch true color LCD display
Instrument size
Length, width and thickness: 196mm x 92mm x 54mm
Jaw size
_7.5 mm *13 mm
Sampling rate
About 3 times / sec.
Range
AC Voltage: 0.00V~600V
AC Current: 0.0mA~20A
Phase: 0.0 ~360.
Frequency: 45.00Hz-65.00Hz
Active power: 0.0W~12kW
Reactive power: 0.0W~12kVAR
Apparent power: 0.0W~12KA
Power factor: - 1 ~+1
Current vector sum: 0mA~60.0A
Resolving power
Voltage: AC 0.01V
Current: AC 0.1 mA
Phase: 0.1 degree
Frequency: 0.01Hz
Active power: 0.1W
Reactive power: 0.1 VAR
Apparent power: 0.1VA
Power factor: 0.001
Current vector sum: 1mA
accuracy
(Baseline conditions)
Voltage: +(0.5% range)
Current: +(0.5% range)
Phase: +1 degree (Note: Phase error under working conditions: 10 mA-20A is +3 degree; below 10 mA is +6 degree)
Active power (+1.0%)
Reactive power (+1.0%)
Apparent power (+ 1.0%)
Frequency (+1.0%)
Power factor (+0.03)
Phase sequence
Positive phase: U1, U2, U3 or I1, I2, I3 cursors flicker sequentially from left to right
Inverse phase: U1, U2, U3 or I1, I2, I3 cursors flicker in turn from right to left
Detection rate
About 2 seconds / time.
Data retention
Press HOLD key to keep data in test, and "HD" symbol display
data storage
500 group
Charging interface
USB Charging Interface
Automatic shutdown
After about 15 minutes, the instrument shuts down automatically to reduce battery consumption.
Backlight function
Yes, suitable for dark places and night use.
Voltage detection
When the battery voltage is less than 3.2V, the low symbol of the battery voltage is displayed to remind the battery to charge.
Instrument quality
Host: 550G (with battery)
Current clamp: 185g x 3
Test line: 250g
Test line length
1.5m
Current clamp length
2m
Working temperature and humidity
- 10 ~40 below 80% Rh
Storage temperature and humidity
- 10 ~60 below 70% Rh
Input impedance
The input impedance of the test voltage is: 1M
Withstand voltage
Sinusoidal AC voltage withstanding 1000V/50Hz between instrument circuit and shell lasts 1 minute
Insulation
Between Instrument Circuit and Sheath Shell (> 100M)
Structure
Dual Insulation with Insulated Vibration Protection Sheath
Suitable for safety regulations
IEC 61010-1 CAT III 600V, IEC 61010-031, IEC 61326, pollution level 2