1. Output test voltage range: AC is 0 ~ 250V; output test voltage accuracy: 50 ~ 250V ± 5% ± 2 words.
2. Leakage current test range: AC / DC, 0 ~ 1MHz, sine wave and composite waveform, three range files are 200μA, 2mA, 10mA.
a. Leakage current test accuracy: AC
200μA file: 5μA ~ 190μA ± 2%;
2mA file: 10μA ~ 1.9mA ± 2%;
10mA file: 0.5mA ~ 10mA ± 2%.
b. Leakage current test accuracy: DC (patient leakage current, patient auxiliary current)
200μA file: 5μA ~ 190μA ± 2%;
2 mA file: 10μA ~ 1.9mA ± 2%.
3. Leakage current preset range and alarm accuracy:
AC: 10μA~10mA ±2%, 200μA/2mA/10mA third gear can set the alarm value arbitrarily.
DC: 10μA ~ 2mA ± 2%, 200μA / 2mA second gear can set the alarm value arbitrarily.
4. Isolation transformer capacity: 500VA standard (optional 300VA ~ 5000VA various specifications)
5. Output test voltage waveform: 50Hz ± 2Hz, sine wave.
6. Timing time range and accuracy: 1S ~ 99S ± 2%, can also be tested manually.
7. Volume: 350 mm × 310 mm × 210 mm.
8. Weight: 14kg.
9. Power supply: 220V ± 10%; AC is 50Hz ± 2Hz.
10. Working conditions
Ambient temperature: 0 to 40 °C.
Relative humidity: no more than 75%.
Atmospheric pressure: 101.25 KPa.